12 Auto zero function

Auto zero is an automatic correction of a possible zero drift.
~ Au0 OFF = auto zero disabled
~ Au0 1
= soft autozero
~ Au0 2
= medium autozero
~ Au0 3
= heavy auto zero
~ Au0 3E
= heavy auto zero over all range
1.
From zero condition on display, press and keep pressed the MENU button until the
acoustic alarm is over, and then release the button. The message "unitS" is
displayed, then press MENU button until the message "AUTO 0" is displayed, then
press PRINT to confirm.
2.
Pressing MENU or CAL button it will be possible to scroll forward or backward
through the different auto zero levels, select the one you wish and confirm it by
pressing the PRINT button.