Intelligent PH Series User's Operation Manual page 14

Precision toploading balances
Table of Contents

Advertisement

5. After having stored the value of internal calibration weight, balance returns to
normal weighing conditions.
6. Return to calibration menu as described at paragraph 6.2 and set the desired
calibration mode: internal, automatic or external.
ATTENTION : this procedure must be effected only using E2 or Class 1 reference
masses.
14