1. Press and keep CAL button pressed with empty pan until the acoustic alarm stops,
then release the button. On display it will be visualized the string "-CAL-", followed by
flashing string "LOAD".
2. L
o
a
d
o
n
t
h
e
p
a
n
a
w
ei
g
ht equal higher or lower than default calibration weight; the balance will recognize it
as valid weight if equal or higher than calibration weight as long as it is a whole
number in comparison with the most meaningful digit of calibration weight.
e.g.: if calibration weight is 200g, it will be possible to calibrate the balance with
values from 100g 200g, 300g, 400g up to the highest limit of balance weighing
range.
The message "LOAD" on display will stop flashing. Once calibration has been
effected, the value of calibrated weight will be displayed.
3. Unload calibration weight. The balance is ready for weighing operations.
NOTE: if
interference
calibration
an error
will be
displayed.
11
there is
during
process,
message