FUSE REPLACEMENT
If the need to check or replace a fuse arises please follow Steps 1 through 4 carefully.
1. Turn scale power switch to the OFF position and unplug power cord from wall outlet.
ALWAYS remove power source from scale before removing or
replacing fuse to prevent risk of electrical shock.
2. Press inward on the fuse holder cap, turn counterclockwise and release. The cap along with the
fuse should then be ejected on the fuse holder. See figure 4.
3. Pull out the old fuse. Place new fuse in holder: 3AG 1/4 ASB type for 110 VAC; 3AG 1/8 ASB
type for 220 VAC. Snap the fuse holder back into position.
4. The power may now be turned on.
To prevent risk of fire, be certain to use type and rating recommended for your Model
scale.
ADJUSTABLE FOOT
OMITTED FOR CLARITY
ON/OFF SWITCH
CAUTION LABEL
The following guide is intended to help isolate and correct minor problems. If, after consulting this
guide, you are unable to correct the problem, contact your Detecto dealer for assistance.
SYMPTOM
System Dead
Incorrect Scale
Weight
Unstable weight
Incorrect rates

TROUBLESHOOTING GUIDE

PROBABLE CAUSE
No AC power
Fuse blown
No program module
Improperly seated
program module
Incorrect software
Improperly calibrated
Scale base not level
Noisy power source
Power not grounded
Vibration and/or
excessive air flow
Improper service
Improper installation
12
POWER CORD
Figure 4
Fuse Replacement
FUSE
PROBABLE SOLUTION
Check power cord
Replace fuse
Insert program module
Re-seat program module
Make sure software is for
same model scale
Calibrate Scale
Make sure scale base is level
Install line filter or conditioner
Ground the power outlet
Relocate scale
Select the proper service
Contact Detecto dealer