Rice Lake 610-10-2 Series Technical & Operating Instructions

Rice Lake 610-10-2 Series Technical & Operating Instructions

Neonatal scale

Advertisement

Rice Lake Neonatal Scale
610-10-2 Series
Technical &Operating
Instructions
To be the best by every measure
118767

Advertisement

Table of Contents
loading

Summary of Contents for Rice Lake 610-10-2 Series

  • Page 1 Rice Lake Neonatal Scale 610-10-2 Series Technical &Operating Instructions To be the best by every measure 118767...
  • Page 3: Table Of Contents

    Technical training seminars are available through Rice Lake Weighing Systems. Course descriptions and dates can be viewed at www.ricelake.com or obtained by calling 715-234-9171 and asking for the training department © 2010 Rice Lake Weighing Systems. All rights reserved. Specifications subject to change without notice. November 2010...
  • Page 4 Rice Lake continually offers web-based video training on a growing selection of product-related topics at no cost. Visit www.ricelake.com/webinars. Neonatal Technical and Operating Instructions...
  • Page 5: Introduction

    This manual can be viewed and downloaded from the Rice Lake Weighing Systems web site at www.ricelake.com/health. Rice Lake Weighing Systems is an ISO 9001 registered company Figure 1-1.
  • Page 6: Assembly Instructions

    Assembly Instructions The Rice Lake Neonatal Scale comes with three pieces. Those pieces include: • Weighing tray • Base assembly • Power supply Remove each part to be assembled from its carton and unwrap the packing material carefully to prevent scratching the unit’s parts.
  • Page 7 The Rice Lake Neonatal Scale is also capable of running on its internal sealed lead-acid rechargeable battery if no additional power source is available. The battery life is approximately 75 hours. If the indicator is LO Bat showing on the display, recharge battery or connect the scale to an AC power source as soon as possible for accurate weighing.
  • Page 8: Scale Setup

    Scale Setup Figure 3-1. Neonatal Display Use the following table to identify what each part of the display does. Item Description Function ON-OFF key This key turns the scale on and off One short push for On One short push for Off REWEIGH key This key allows repeated reweighing of an One short push...
  • Page 9 Use the following steps to set up your Rice Lake Neonatal Scale when you initially power it up. 1. To operate the unit using the AC adaptor, plug the AC adaptor cable into the back of the scale as shown in Figure 2-2 2, and the AC adaptor plugs into a power source.
  • Page 10: Scale Operation

    Scale Operation The Rice Lake Neonatal Scale can accurately weigh neonates and babies up to 30 lbs. Use the following steps to weigh a baby. 1. Press the key to turn the scale on. ON-OFF 2. Wait until the (if set up in Lb/Oz) appears on the digital display or if set up in Kg.
  • Page 11: Calibration

    Calibration The Rice Lake Neonatal Scale is pre-calibrated from the factory. But if you need to calibrate the scale, the scale must be turned on. Once the scale is turned on, press the keys simultaneously. Hold those REWEIGH ZERO keys until appears on the display alternating with the suggested weight for calibration.
  • Page 12: Troubleshooting & Testing

    Troubleshooting & Testing Refer to the following instructions to check and correct any failure before consulting service personnel. Symptom Possible Cause Corrective Action Scale does not turn on when using the Dead battery Connect the scale to a power source rechargeable battery Scale does not turn on Dead battery...
  • Page 13: Test Mode

    Test Mode The test mode menu is a special mode used for checking the systems’ state and for troubleshooting. Entry into this mode is possible only when the scale is turned off. To access the test mode parameter, press and hold the key when starting the unit.
  • Page 14: Maintenance

    Maintenance The following section provides instructions for maintaining and cleaning the Rice Lake Neonatal Scale. Maintenance operations other than those described in this section should be performed by qualified service personnel. Basic Maintenance Before the first use of the scale and after periods of non-use, check the scale for proper operation and function. If the scale does not operate correctly, contact qualified service personnel.
  • Page 15: Specifications

    Specifications Capacity and Graduation 0 - 5 Lb x 0.05 oz (0 - 2 Kg x 1 g) 5 - 12 Lb x 0.1 oz (2 - 6 Kg x 5 g) 12 - 30 Lb x 0.2 oz (6 - 15 Kg x 5 g) Power Requirements: 120 VAC-9VDC-50Hz / 230 VAC-9VDC-50Hz Environmental:...
  • Page 16: For More Information

    Fax Number 715-234-6967 Email • US sales and product information at prodinfo@ricelake.com • International (non-US) sales and product information at intlsales@ricelake.com Mailing Address Rice Lake Weighing Systems 230 West Coleman Street Rice Lake, WI 54868 USA Neonatal Technical and Operating Instructions...
  • Page 17: Neonatal Scale Limited Warranty

    Neonatal Scale Limited Warranty Rice Lake Weighing Systems (RLWS) warrants that all RLWS equipment and systems properly installed by a Distributor or Original Equipment Manufacturer (OEM) will operate per written specifications as confirmed by the Distributor/OEM and accepted by RLWS. All systems and components are warranted against defects in materials and workmanship for two years.
  • Page 18 PN 118767 11/10...