Controller Parameters

Home Display
°C: measured temperature in Celsius. Temperature units
cannot be changed without entering the configuration.
Contact Customer Service if a different temperature unit
is required.
OP: % output power demand; displayed in lower display
(cannot be changed).
C.id: Controller identification number.
IdHi: Deviation High Alarm
tunE: One-shot autotune enable.
run LiSt (Program Run List)
StAt: Displays the program status [OFF, run (running
active program), hoLd (program on hold), HbAc (waiting
for process to catch up), End (program completed)] in the
lower display. The controller will default to "OFF."
FASt: Fast run through program (no/YES). The controller
will default to "no."
SEG.d: Flash active segment type in the lower display of
the home display (no/YES). The controller will default to
"no."
ProG LiSt (Program Edit List)
Hb: Press the UP or DOWN ARROW to select the hold-
back type [OFF (disables holdback), Lo (deviation low
holdback), Hi (deviation high holdback) or bAnd (deviation
band holdback)] for the entire program. The controller will
default to "OFF."
Hb.U: Press the UP or DOWN ARROW to select the
holdback value (in display units).
rmP.U: Press the UP or DOWN ARROW to toggle
between ramp units (SEc, min or Hour). Controller will
default to "SEc."
dwL.U: Press the UP or DOWN ARROW to toggle
between dwell units (SEc, min or Hour). Controller will
default to "SEc."