Barnstead F62730 Operation Manual page 27

Furnace
Table of Contents

Advertisement

8 S
EGMENT
rmP.U: Setting Ramp Units
Ramp units are time units which are used in "rmP.r" seg-
ments (ramp to a setpoint at degrees per second, minute
or hour) and "rmP.t" segments (ramp to setpoint in a spe-
cific amount of time). See "Setting the Segment Type" for
an explanation on how to set a ramp segment.
1.
Press the PAGE button until you reach the pro-
gram list (ProG LiSt).
2.
Press the SCROLL button until display reads,
"rmP.U."
3.
Press the UP or DOWN button to toggle
between seconds, minutes and hours.
dwL.U: Setting Dwell Units
Dwell units are time units which are used in "dwELL" seg-
ments (amount of time to remain at a specific tempera-
ture). See "Setting the Segment Type" for an explanation
on how to set a dwell segment.
1.
Press the PAGE button until you reach the pro-
gram list (ProG LiSt).
2.
Press the SCROLL button until display reads,
"dwL.U."
3.
Press the UP or DOWN button to toggle
between seconds, minutes and hours.
CYC.n: Setting the Number of Cycles
Set the number of times a group of segments or pro-
grams are to be repeated by following the steps listed
below.
1.
Press the PAGE button until you reach the pro-
gram list (ProG LiSt).
2.
Press the SCROLL button until display reads,
"CYC.n."
3.
Press the UP or DOWN button to select the
number of cycles you want to run or, press the
DOWN button to select "cont." so the program
will run continuously.
P
M
ROGRAMMABLE
ODEL W
/OTP
27