8 S
EGMENT
4.
Press the SCROLL button until display reads,
"tGt."
5.
Press the UP or DOWN button to set a target
setpoint.
6.
Press the SCROLL button until display reads,
"dur."
7.
Press the UP or DOWN button to select a time
in ramp units (seconds, minutes or hours; set in
the "rmP.U" parameter.
dwEll
To maintain a constant temperature for a specified time:
1.
Press the PAGE button until you reach the pro-
gram list (ProG LiSt).
2.
Press the SCROLL button until display reads,
"tYPE."
3.
Press the UP or DOWN button until display
reads, "dwEll."
4.
Press the SCROLL button until display reads,
"dur."
5.
Press the UP or DOWN button to select a time
in dwell units (seconds, minutes or hours; set in
the "dwL.U" parameter).
StEP
To climb instantaneously from the current temperature to
a specified temperature.
1.
Press the PAGE button until you reach the pro-
gram list (ProG LiSt).
2.
Press the SCROLL button until display reads,
"tYPE."
3.
Press the UP or DOWN button until the display
reads, "StEP."
4.
Press the SCROLL button until display reads,
"tGt."
5.
Press the UP or DOWN button to set a target
setpoint.
P
M
ROGRAMMABLE
ODEL W
/OTP
29