8 S
P
M
/OTP
EGMENT
ROGRAMMABLE
ODEL W
Cyc.n: Press the UP or DOWN ARROW to set the num-
ber of program cycles (1 to 999 or cont). The controller
will default to "cont."
SEG.n: Press the UP or DOWN ARROW to select the
segment number (1-8 in 8 segment models).
tYPE: Press the UP or DOWN ARROW to select the seg-
ment type [End (end of program), rmP.r = ramp rate (ramp
to a specified setpoint at a set rate), rmp.t = ramp time
(ramp to a specified temperature in a set time), dwEll (to
maintain a constant temperature for a set time), StEP
(climb instantaneously from current to specified tempera-
ture). The controller will default to "End." Other parame-
ters used with tYPE include; tGt target setpoint), Rate
(rate of temperature increase) and dur (time to target set-
point or time to dwell)].
End.t: End segment type: dwELL (dwell continuous), rSEt
(reset) and S OP (End Segment Output power level.
AL LiSt (Alarm List)
IdHi: Deviation High Alarm.
Atun LiSt: (Autotune List)
tunE: One-shot autotune enable.
drA: Adaptive tune enable.
drA.t: Adaptive tune trigger level in display units. Range =
1—9999.
Pid LiSt
G.SP (Gain Setpoint): Is the temperature at which the
controller switches from the (SEt 1) PID values to the
(SEt 2) PID values.
Pb: Proportional band in display units. (SEt 1)
ti: Integral time in seconds. (SEt 1)
td: Derivative time in seconds. (SEt 1)
Pb2: Proportional band. (SEt 2)
ti2: Integral time in seconds. (SEt 2)
td2: Derivative time in seconds. (SEt 2)