SECTION 6—OPERATING INSTRUCTIONS

SensO

Oxygen Purity Indicator - Models
2
IRC5PO2AW & IRC5PO2VAW
This feature monitors the purity of oxygen generated by the 
oxygen concentrator. If purity falls below factory preset 
standards, indicator lights on the control panel will 
illuminate.

Initial Startup of the Concentrator

NOTE: Concentrator may be used during the initial start warm‐up 
time (approximately 30 min.) while waiting for the O
 purity to 
2
reach maximum.
When the unit is turned on, the GREEN light will come on 
(SYSTEM OK/O
 greater than 85%). After five minutes, the 
2
oxygen sensor will be operating normally and will control the 
indicator lights depending on oxygen concentration values. 
Explanation of Oxygen Purity Indicator Lights -
Model IRC5PO2AW & IRC5PO2VAW
NOTE: For this procedure, refer to FIGURE 6.6 on page 23.
GREEN light (O
) ‐ Normal Operation.
2
YELLOW light (
) ‐ Immediately call supplier. You may 
continue to use the concentrator unless instructed otherwise 
by your supplier. Be certain that backup oxygen is nearby.
RED light ( ) ‐ Total unit shutdown. Immediately switch to 
backup oxygen supply and call supplier.
GREEN light ‐ with YELLOW light flashing ‐ Immediately call 
supplier. Oxygen sensor malfunctioning; you may continue to 
use the concentrator. 
Perfecto
™Series
22
Part No 1160877
2