Section 5-Typical Product Parameters - Invacare Perfecto2 IRC5PO2AW User Manual

Perfecto2 series oxygen concentrators without senso2 | oxygen concentrators with senso2 | homefill compatible
Hide thumbs Also See for Perfecto2 IRC5PO2AW:
Table of Contents

Advertisement

SECTION 5—TYPICAL PRODUCT
PARAMETERS
IPX1
Electrical
Requirements:
Rated Current Input:
Sound Level:
Altitude:
Oxygen Output
Concentration Levels:
NOTE: Stated 
concentration levels 
achieved after initial 
warm‐up period 
(approximately 30 
minutes).
Part No 1160877
SECTION 5—TYPICAL PRODUCT PARAMETERS
Alternating Current
Type BF equipment
Unit running
Unit not running
Attention - Consider Accompanying Documents
DO NOT smoke
Class II, Double Insulated
Protected against dripping water
230 VAC ± 10% (253 VAC/207 VAC), 50 Hz
1.4 A
39.5 dBA Average
All Perfecto
Models: Up to 8,000 ft (2438 meters) above sea
2
level without degradation of concentration levels. From 8,000 ft
to 13,129 (4000 meters) below 90% efficiency.
All Perfecto
Models while filling a HomeFill System:
2
Up to 6,000 ft (1828 meters) above sea level without degrada-
tion of concentration levels. From 6,000 ft (1828 meters) to
13,129 ft (4000 meters) below 90% efficiency
87% to 95.6% at 0.5 to 5 L/min
93% ± 3% at 2 L/min
13
Perfecto
™Series
2

Advertisement

Table of Contents
loading

This manual is also suitable for:

Perfecto2 irc5po2vawPerfecto2 irc5paw

Table of Contents