SECTION 5—TYPICAL PRODUCT PARAMETERS
IRC5PO2AW & IRC5PO2VAW INDICATORS
LABEL
SYMBOL
O
2
LABEL
SYMBOL
Perfecto
™Series
2
O
PURITY
2
SYSTEM OKAY
O
over 85% (± 2%)
2
O
Between 73% (± 3%)
2
to 85% (±2%)
SYSTEM FAILURE
O
Below 73% (±3%)
2
IRC5PAW INDICATORS
STATUS
SYSTEM OKAY
SYSTEM FAILURE

INDICATOR LIGHTS

GREEN Indicator Light
YELLOW Indicator light
A. YELLOW Solid
B. YELLOW Flashing Sensor
Failure
Call a qualified technician.
RED Indicator Light
Continuous Audible Alarm
Sieve-GARD Compressor
Shutdown
INDICATOR LIGHTS
GREEN Indicator Light
RED Indicator light
Continuous Audible Alarm
Sieve-GARD™Compressor
Shutdown
Call a qualified technician.