Invacare Platinum IRC5LXAW Operator's Manual page 17

Oxygen concentrators standard, with/senso2 and homefill
Hide thumbs Also See for Platinum IRC5LXAW:
Table of Contents

Advertisement

Regulatory Listing:
CE marked models:
Electrical:
Placement:
Tubing:
Relative Humidity:
Time of Operation:
Recommended Storage and
Shipping Temperature:
Part No 1134866
SECTION 4—TYPICAL PRODUCT PARAMETERS
IRC5LXAW/IRC5LXO2AW/
IRC5LXO2AWQ/IRC9LXO2AWQ
ETL tested as complying with:
EN55011
EN61000-3-2
EN61000-3-3
IEC 60601-1, A1, A2
IEC 60601-1-2
IRC5LXO2AW, IRC5LXO2AWQ,
IRC9LXO2AWQ Models Only
ISO8359
MDD
IRC5LXO2AW, IRC5LXO2AWQ,
IRC9LXO2AWQ
No extension cords.
No closer than 7.5 cm (3 inches) from any
wall, furniture, draperies, or similar surfaces.
IRC5LXAW, IRC5LXO2AW,
IRC5LXO2AWQ
2.1 m (7 ft) cannula with a maximum 15.2 m
(50 ft) of Crush-Proof Tubing (DO NOT
pinch).
IRC9LXO2AWQ
Recommended use up to 15.2 M (50 ft) high
flow tubing with high flow cannula at all flow
rates.
20 to 60%
Up to 24 hours per day.
-29°C to 65°C (-20°F to 150°F) at 15-95%
relative humidity.
17
Platinum™

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents