Invacare Platinum IRC5LXAW Operator's Manual page 15

Oxygen concentrators standard, with/senso2 and homefill
Hide thumbs Also See for Platinum IRC5LXAW:
Table of Contents

Advertisement

*Oxygen Output
Concentration Levels:
All 5LXO2AWQ/5LXAW/
5LXO2AW/9LXO2AWQ
models
* (Stated concentration levels
achieved after initial warm-up
period (approximately 30
minutes).
Maximum Outlet Pressure:
Flow Range:
Low Flow Alarm:
Average Power Consumption: (IRC5LXO2AWQ)
Pressure Relief Mechanism
Operational at:
Part No 1134866
SECTION 4—TYPICAL PRODUCT PARAMETERS
IRC5LXAW, IRC5LXO2AW and
IRC5LXO2AWQ
93% minimum at 1 to 3 L/min.
91% minimum at 4 L/min.
87% minimum at 5 L/min. (maximum
recommended flow)
IRC9LXO2AWQ
92% minimum at 1 to 2 L/min.
91% minimum at 8 L/min.
87% minimum at 9 L/min.
93% minimum at 3 to 7 L/min.
IRC5LXAW, IRC5LXO2AW,
IRC5LXO2AWQ
34.5 ± 3.45 kPa (5 ± 0.5 psi)
IRC9LXO2AWQ
62.0 kPa ± 3.45 kPa (9 ± 0.5 psi)
IRC5LXAW, IRC5LXO2AW,
IRC5LXO2AWQ
0.5 to 5 L/min. Flowrates less than 1 L/min.
are not recommend
IRC9LXO2AWQ
1 L/min. to 9 L/min. Flowrates less than 1
L/min. are not recommended.
0 L/min. to 0.5 L/min. Rapid Audible Alarm
Beeping (No Accessories Connected).
300 W
280 W @ 3 L/min.
(IRC5LXO2AW)
365 W
340 W @ 3 L/min.
380 W (IRC5LXAW)
475 W (IRC9LXO2AWQ)
241 kPa ± 24.1 kPa (35 psi ± 3.5 psi)
15
Platinum™

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents