Invacare Platinum IRC5LXAW Operator's Manual page 16

Oxygen concentrators standard, with/senso2 and homefill
Hide thumbs Also See for Platinum IRC5LXAW:
Table of Contents

Advertisement

SECTION 4—TYPICAL PRODUCT PARAMETERS
Change in maximum
recommended flow when
back pressure of 7k Pa is
applied:
Filters:
Safety System:
Width:
Height:
Depth:
Weight:
Shipping Weight:
Operating Ambient
Temperature:
Exhaust:
Oxygen Output:
Cabinet:
Platinum™
0.7 L/min.
Cabinet (2), Outlet HEPA and Compressor
Inlet.
Current overload or line surge shutdown.
High temperature compressor shutdown.
High Pressure Alarm w/compressor
shutdown.
Low Pressure Alarm w/compressor
shutdown.
Battery Free Power Loss Alarm.
SensO
Oxygen System (SensO
2
Flow Alarm.
46.7 cm ± 1 cm (18 3/8 in ± 3/8 in)
67.0 cm ± 1 cm (6 3/8 in. ± 3/8 in)
36.5 cm ± 1 cm (14 3/8 in. ± 3/8 in)
IRC5LXAW, IRC5LXO2AW,
IRC5LXO2AWQ
23.6 kg ± 1 kg (52 lbs ± 2 lbs)
IRC9LXO2AWQ
24.0 kg ± 1 kg (53 lbs ± 2 lbs)
IRC5LXAW, IRC5LXO2AW,
IRC5LXO2AWQ
26.8 kg ± 1 kg (59 lbs ± 2 lbs)
IRC9LXO2AWQ
27.2 kg ± 1 kg (60 lbs ± 2 lbs)
10°C - 35°C (50°F - 95°F) at 20-60% relative
humidity.
Less than Ambient +19°C (+ 35°F)
Less than Ambient +2°C (+ 4°F).
Impact Resistant flame-retardant plastic
cabinet that conforms to UL 94-V0.
16
Model) Low
2
Part No 1134866

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents