Rice Lake Summit SB-1000 Installation And Operation Manual

Rice Lake Summit SB-1000 Installation And Operation Manual

Bariatric wheelchair scale

Advertisement

Quick Links

Summit SB-1000
®
Bariatric Wheelchair Scale
Installation & Operation
Manual
106992

Advertisement

Table of Contents
loading

Summary of Contents for Rice Lake Summit SB-1000

  • Page 1 Summit SB-1000 ® Bariatric Wheelchair Scale Installation & Operation Manual 106992...
  • Page 3: Table Of Contents

    Course descriptions and dates can be viewed at www.ricelake.com or obtained by calling 715-234-9171 and asking for the training department © 2009 Rice Lake Weighing Systems. All rights reserved. Printed in the United States of America. Specifications subject to change without notice. October 2009...
  • Page 4 Summit SB-1000 Bariatric Wheelchair Scale Installation Manual...
  • Page 5: Introduction And System Overview

    The Summit SB-1000 Bariatric Wheelchair Scale is a fully electronic, low profile floor scale that measures 36 in x 36 in (.91 m x .91 m), and has a capacity up to 1000 lbs (500 kg). The SB-1000 uses four corner-mounted, alloy steel shear beam load cells, with the cells recessed into the frame channels for protection.
  • Page 6: Scale Installation

    10-ft length of load cell cable. The Summit SB-1000 Bariatric Wheelchair Scale has one threaded eyebolt hole located in the center of the deck to allow installation of an eyebolt hook for use when lifting the scale with chains. It’s recommended to use a 1/2 in-20NF eyebolt to lift the scale.
  • Page 7: Bariatric Wheelchair Scale Feet Assembly And Adjustment

    NOTE:The cable is pre-installed from the factory but should replacement be required in the future, cable installation instructions are shown below. The junction box is easily accessible through an access plate located on the side of the Summit SB-1000. Use the following steps to wire up the junction box.
  • Page 8: Access Ramps

    11. Secure the Summit SB-1000 side plate. Access Ramps Access ramps for Summit SB-1000 Bariatric Wheelchair Scales are designed to bolt to the floor, with built-in mounting plates that attach to the scale feet. When used with access ramps, side movement of the scale is automatically eliminated, and no other mounting plates are necessary.
  • Page 9: Adjustments And Calibration

    Adjustments and Calibration The following sections describe adjustments that need to be made to the Summit SB-1000 Bariatric Wheelchair Scale. Mechanical Adjustments To accommodate minor floor unevenness, the scale feet can be used to adjust scale height up or down a fraction of an inch.
  • Page 10: Calibration Procedure

    70% - 80% of the scale’s capacity. If several weights are used, they should be evenly distributed around the platform. Perform a span calibration. Remove the test weights and check the zero reading. Repeat the calibration process if necessary. Summit SB-1000 Bariatric Wheelchair Scale Installation Manual...
  • Page 11: Indicator Installation And Configuration

    AC Power Connection The Summit SB-1000 Scale has a 120 VAC adaptor or 230 VAC adaptor to use when power is readily available. The AC power adaptor plugs into the back of the indicator as shown in Figure 4-1. The AC adaptor, when not in use, plugs into the back housing of the indicator.
  • Page 12: Scale Configuration

    After configuration or calibration is done, the jumper must be replaced for normal scale operation. NOTE: A display of Con En indicates that the jumper is not in place. Put jumper on both pins to return the scale to normal weigh mode. Summit SB-1000 Bariatric Wheelchair Scale Installation Manual...
  • Page 13: Numeric Data Entry

    4.3.1 Configuration Mode To get into the configuration mode, turn the scale off and remove the configuration jumper as shown in Figure 4-3. Turn the scale on. While is displayed, press and hold the key until appears on the Start Kg-Lb IDENT display.
  • Page 14: Programming Mode

    REWEIGH key. Weight algorithm A t INE <-> 10 This parameter is preset from the maximal exponent factory. To advance to the next step, press the REWEIGH key. Table 4-2. Configuration Mode Menu Summit SB-1000 Bariatric Wheelchair Scale Installation Manual...
  • Page 15 Step Function Display Available Parameters Message style on MESS <-> WEIGH This displays the message that will weight algorithm show on the indicator display. To change the message, press the Kg-Lb key. WEIGH, LIVE, - - - - - - To advance to the next step, press the REWEIGH key.
  • Page 16 Double click on menu. the Kg-Lb key to change the options value or to advance to the next step, press the REWEIGH key. 0= enable 1= disable Table 4-2. Configuration Mode Menu Summit SB-1000 Bariatric Wheelchair Scale Installation Manual...
  • Page 17: Reset To Factory Defaults

    Step Function Display Available Parameters Set Option defaults Set < - > 1 To change, press the Kg-Lb key. 0 - default Options = (OPx) 1 = US default (US defaults are OP1-1, OP2-0, OP3-0, OP4-0, OP5-0, OP6-1, OP-7-0, OP8-1, OP9-1, OP10-1).
  • Page 18: Scale Calibration

    ZERO key. dONE The scale displays that it has saved that calibration value and automatically advances to the last step. StArt The scale reboots. Con En Replace the configuration jumper. Table 5-1. Calibration Menu Summit SB-1000 Bariatric Wheelchair Scale Installation Manual...
  • Page 19: Scale Operation

    Scale Operation The scales have the capability of performing different operations beyond just calculating weight. The various operating instructions are described below. Weighing Use the following steps to weigh a person. 1. Press the key to turn on the scale and will appear on the display.
  • Page 20: Using The Tare Function

    5. You can now accurately weigh the patient. 6. To cancel the tare weight, press and hold the key until disappears from the display. The tare TARE TARE weight is also cancelled when the scale is turned off. Summit SB-1000 Bariatric Wheelchair Scale Installation Manual...
  • Page 21: Rs-232 Communications

    The scale comes with an RS-232 port which enables weight data to be transmitted to other equipment, such as a computer or printer. The RS-232 cable with DB-9 connector (PN 100719) is available from Rice Lake Weighing System. Figure 4-1 on page 7 shows where the RS-232 connection is.
  • Page 22 • Unit of measurement (metric or constitutional) (UOM) = c (m or c) Table 7-1. RS-232 Communication Parameters If you’re using the Rice Lake files, please follow the instructions below. 1. Go to www.ricelake.com/health and download the Rswin.exe and Inbar.ini files located in the downloads section of the web site and download them to your computer.
  • Page 23 3. Double click on the Rswin.exe file and the following screen appears. C l i c k o n F I L E S and in dropdown m e n u , s e l e c t L O A D CONFIGURATION as noted in Step 4 Figure 7-1.
  • Page 24: Sample And Explanation Of Esc Protocol

    Figure 7-4. ASCII Screen Example Sample and Explanation of ESC Protocol When the scale measures weight and sends this over the communications line to the PC, the string will look like this. <SCALE> --------<PC> Summit SB-1000 Bariatric Wheelchair Scale Installation Manual...
  • Page 25 When the user wants to diagnose any problems with the scale, the operator will have to ask the scale to send the error data (if any exists). This is done with the Diagnostics (request) command and will look like this: If the battery is okay, the scale will reply with the following value: If the battery is critically low, it will reply with: RS-232 Communications...
  • Page 26: Service Information

    Use simulator or test indicator for stability. Service indicator Display stays at zero Indicator faulty Service indicator Load cell connections are faulty Check the cable connections in junction box and at the indicator Table 8-1. Troubleshooting Table Summit SB-1000 Bariatric Wheelchair Scale Installation Manual...
  • Page 27: Test Mode

    Problem Symptom Description Solution The display shows a The load on the scale exceeds the Remove the excess weight and use the scale STOP message capacity of the scale according to manufacture’s specs. The display shows The battery is low Recharge the battery.
  • Page 28: Periodic Maintenance

    Use care with high pressure steam washdowns for hermetically-sealed load cells. The steam may not damage the load cells, but the elevated temperatures may cause incorrect readings until the unit cools to room temperature. Summit SB-1000 Bariatric Wheelchair Scale Installation Manual...
  • Page 29: Load Cell Replacement

    Load Cell Replacement Replacement load cells can be ordered from Rice Lake Weighing Systems. To replace a load cell, lift the scale with a chain and remove foot, then remove the defective load cell. Disconnect load cell cable from the junction box and cut cable ties.
  • Page 30: Load Cell Wiring To Junction Box

    Corner correction trimming and calibration is necessary after load cell replacement. Follow instruction in Sections 3.2 on page 5. Cable Color Code J-Box Terminal +Excitation Black -Excitation Green +Signal White -Signal Table 8-3. Load Cell Wiring Summit SB-1000 Bariatric Wheelchair Scale Installation Manual...
  • Page 31: Bariatric Wheelchair Scale Specifications

    Bariatric Wheelchair Scale Specifications Electrical Grounding For systems where the scale is connected to a 115 VAC circuit, the indicator must be directly connected to an earth ground with a ground interface cable of no more than 3W resistance throughout its length. Load Cell Excitation Rated Excitation: 10 VDC Maximum Excitation: 15 VDC...
  • Page 32: Summit Sb-1000 Limited Warranty

    Summit SB-1000 Limited Warranty Rice Lake Weighing Systems (RLWS) warrants that all RLWS equipment and systems properly installed by a Distributor or Original Equipment Manufacturer (OEM) will operate per written specifications as confirmed by the Distributor/OEM and accepted by RLWS. All systems and components are warranted against defects in materials and workmanship for two years.