Summit SB-1150 ™ Bariatric Wheelchair Scale Software Version 11525 Technical Manual August 17, 2022 PN 213454 Rev A...
Page 2
All information contained within this publication is, to the best of our knowledge, complete and accurate at the time of publication. Rice Lake Weighing Systems reserves the right to make changes to the technology, features, specifications and design of the equipment without notice.
Page 3
Initial manual release; formatted content to match other medical manuals; software version 11525 Table i. Revision Letter History Technical training seminars are available through Rice Lake Weighing Systems. Course descriptions and dates can be viewed at www.ricelake.com/training or obtained by calling 715-234-9171 and asking for the training department.
7.0 Specifications ............... . . 23 Rice Lake continually offers web-based video training on a growing selection of product-related topics at no cost.
Introduction The Summit SB-1150 bariatric wheelchair scale is a heavy-duty scale that offers exceptional performance in clinics and hospitals. Patients walk or can be wheeled onto the weighing platform to be weighed. The scale is a fully electronic, low profile floor scale that measures 36 in x 36 in (.91 m x .91 m) and has a capacity up to 1000 lb (500 kg).
Do not operate or work on this equipment unless this manual has been read and all instructions are understood. Failure to follow the instructions or heed the warnings could result in injury or death. Contact any Rice Lake Weighing Systems dealer for replacement manuals.
Unpacking Remove all packing material and inspect the contents for damage possibly caused during shipment. Contact Rice Lake Weighing Systems and the shipper immediately if there is damage to the scale. The shipping container should contain the scale, scale feet, this manual, the indicator, and a 10 foot length of load cell cable.
Summit SB-1150 Wheelchair Scale Assembly The scale feet are shipped detached from the scale for load cell protection during shipment. 1. Unpack all parts. 2. Ensure each foot is screwed in until the foot touches either the load cell or the underside of the deck.
Power Connection An optional AC power adapter can be used when a power outlet is available. Only use power adapters supplied by or purchased from Rice Lake Weighing Systems. The use of a power IMPORTANT adapter not from Rice Lake Weighing Systems voids the warranty.
Summit SB-1150 Wheelchair Scale Load Cell Connections The indicator and scale comes factory installed with a load cell cable connection. Follow the procedure below if the load cell cable needs to be replaced or reconnected to the indicator. 1. Unscrew and remove the tilt stand bracket from the indicator to gain access to the load cell connection.
Summit SB-1150 Wheelchair Scale Weighing Use the following steps to weigh. to turn on the scale. 0.0 appears on the display along with the ZERO annunciator. 1. Press 2. Place the patient on the scale. The patient’s weight is displayed, the LOCK annunciator is on and the indicator beeps to indicate the end of the weighing process.
Summit SB-1150 Wheelchair Scale Troubleshooting Refer to the following table to check and correct any failure before contacting service personnel. Symptom Possible Cause Corrective Action Scale does not turn on Dead batteries Replace batteries or connect to AC power Faulty electrical outlet...
SB-1150 Bariatric Wheelchair Scale Configuration Mode Use the following steps to enter into Configuration mode. 1. Ensure the scale is turned off. . Continue to hold both keys until Id appears. 2. Turn the scale on by simultaneously pressing ENTER The unit cycles through its startup function and continues to display the software version.
Page 18
SB-1150 Bariatric Wheelchair Scale Parameter Description Choices Steps ASTART Weight process start limit — Maximum (full Value (2.0) The display toggles between a numeric value and ASTART; capacity)/10; Determine when the weight If you do not want to change this value, press the BMI key to move to the algorithm starts (when the “----”...
SB-1150 Bariatric Wheelchair Scale Default Menu The default menu is used to return the scale back to its factory settings and is shown below. DONE PROG CALIB TEST Figure 4-4. Default Menu Use the following steps to return the settings back to their factory default. 1.
The unit comes with an RS-232 port that enables weight data to be transmitted to other equipment, such as a computer or printer. The RS-232 cable with DB-9 connector (PN 100719) is available from Rice Lake Weighing Systems. That connection is shown in USB Connection section.
Summit SB-1150 Wheelchair Scale Samples of Escape Protocol Examples of what is sent to the computer from the scale. direction of communication <SCALE> <PC> <SCALE> <PC> <ESC> R <ESC> W0210.9 <ESC> Nm <ESC> E <ESC>R <ESC>E Reading Reading Unit of Measure...
Summit SB-1150 Wheelchair Scale Maintenance The following section provides instructions for maintaining and cleaning the unit. Do not immerse the scale in cleaning or other liquid solutions. IMPORTANT Do not use Isopropyl alcohol or other solutions to clean the indicator display surface.