Table 3-1 Electrical specifications (continued)
Parameter
Stability (output drift)
Following a 30-minute warm-up, with the output in the
ON state according to the operating mode (CC with load
or CV), and with a change in the output over 8 hours
under constant load, line, and ambient temperature.
Programming accuracy (23 °C ± 5 °C)
Meter readback accuracy (23 °C ± 5 °C)
Programming/Meter resolution
Maximum output float voltage

Physical Characteristics

Table 3-2 Physical characteristics

Parameter
Display
Rotary knob for reading adjustment
Size
Dimensions
Weight
U8030A Series User's Guide
U8031A
• Voltage: <0.02%
• Current: <0.1%
• CV: ≤0.25% + 15 mV
• CC: ≤0.30% + 15 mA
• CV: ≤0.25% + 10 mV
• CC: ≤0.25% + 10 mA
• Voltage: 10 mV (4 digits)
• Current: 10 mA (3 digits)
±240 V
DC
U8031A/U8032A
LCD with amber backlight
Yes
4U, half rack
179.0 mm × 212.3 mm × 379.0 mm
8.2 kg
Characteristics and Specifications

Physical Characteristics

U8032A
3
67