18
GETTING YOUR InDuo™ SYSTEM READY FOR USE
b
(Example)
Apply control solution
• Shake the control solution vial well.
• Remove the cap.
• Squeeze the vial and discard the first drop and
wipe off the dispenser tip to obtain an accurate
result.
• Squeeze the vial again to get a small hanging drop.
• Apply the hanging drop b to the top edge area of
the OneTouch
the narrow channel) until the drop is drawn into
the OneTouch
window is filled and the meter display begins to
count down.
Read test results
c
• The meter will count down from N to A and the
control solution test result will appear c.
• Compare the result with the range printed on the
OneTouch
should fall within this range.
Ultra™ test strip (where it meets
®
Ultra™ test strip, the confirmation
®
®
Ultra™ test strip vial label. The result