Bayer HealthCare Contour next EZ User Manual page 6

Blood glucose monitoring system
Hide thumbs Also See for Contour next EZ:
Table of Contents

Advertisement

• Do not press the test strip against your finger
when testing. This may block the sample tip.
• Let the meter and test strips adjust to the
temperature where you are testing. Whenever the
meter is moved from one location to another, allow
approximately 20 minutes for the meter to adjust to
the temperature of a new location before performing
a blood glucose test. The meter has been designed
to give accurate results at temperatures between
5°C and 45°C.
• Check the expiry dates on your test strips and
control solution. It is important to not use the test
strips or control solution if the expiry date printed on
the bottle label and carton has passed. The control
solution should not be used if it has been six months
since you first opened the bottle. It will help to
write the six month discard date on the control
solution label.
®
6
Information For Safety
• Have all the materials you will need ready
before you begin testing. This includes your
®
C
N
EZ meter, the C
ONTOUR
EXT
strips, the M
®
2 lancing device and M
ICROLET
lancets. You will also need C
solution to run a quality control check.
C
 N
control solutions are sold separately.
ONTOUR
EXT
• Examine product for missing, damaged, or broken
parts. If the test strip bottle is open inside a new
box of strips, do not use those strips. Contact
Bayer Diabetes Care Helpline at 1-800-268-7200 for
replacement parts.
®
• The M
2 lancing device is intended for self
ICROLET
testing by a single patient. It must not be used on
more than one person due to the risk of infection.
®
• Use a new M
lancet each time you test
ICROLET
because it is no longer sterile after use.
• Do not perform a blood glucose test when the
®
C
N
EZ meter is connected to an external
ONTOUR
EXT
device, e.g., computer.
• Use only approved equipment from the manufacturer
or certified body such as UL or TUV.
Need Help? 1-800-268-7200
www.bayerdiabetes.ca
®
N
test
ONTOUR
EXT
®
ICROLET
®
N
control
ONTOUR
EXT
7