Note
The 3.00 A AC minimum specified reading is calculated by
the Meter as 5% of 60 A. However for safety and
compactness, the current shunt in the Meter is specified for
10 A continuous, 20 A overload for up to 30 seconds.
Exceeding these levels will open the user-replaceable
internal fuse.
The 10 A AC current in the 111 and 112 is designed for
accurate, in-line measurement of loads on line-powered
devices and circuits. Because the lowest specified AC
readings are 3.00 A AC and 300 mV AC, the Meter is not
specified for use with AC current clamp accessories.
Measuring Resistance
Measuring Capacitance
True RMS Multimeters
Making Basic Measurements
aej04f.eps
aej05f.eps
7