Acquiring an Oblique Image

Before acquiring an oblique image, check that you have:
Reset the unit rotative arm to the start position for patient to enter the unit.
Selected the patient record.
Accessed the Imaging Window.
Accessed the Cephalometric Acquisition Window.

Preparing the Unit and Setting the Acquisition Parameters

To acquire an oblique image, follow these steps:
1. Raise and block the sensor manually.
2. On the x-ray source assembly, set the collimator selector to AP/PA.
3. Position the head clamps manually for the oblique exam with the desired angle.
4. In the Cephalometric Acquisition Window, click the Program button to access the
Program pane. In the Program pane:
The
for an oblique exam is active.
Click
5. Click the Patient button to access the Patient pane. Select the patient type.
6. Click the Patient button to access the Patient pane. Select the patient:
Type
Size
7. If the default parameter setting is not adapted to your patient type, click the
Parameter button and in the Parameter pane select the appropriate parameters. To
save the new parameter settings as the default settings, click
Memorize settings.
KODAK 8000C Digital Panoramic and Cephalometric Extraoral Imaging System User Guide (SM735)_Ed 02
IMPORTANT
You must position the head clamps manually because they
are not positioned automatically from the Program pane exam
type selection. In this case, the relevant exam type selection
icon becomes active.
to select the desired angle.
Acquiring an Oblique Image
fand select
5–9