Due-Lock Enabled - Honeywell GasAlertMicro 5 User Manual

Gas detector
Table of Contents

Advertisement

GasAlertMicro 5/PID/IR
User Manual
9. Unless disabled in user options, the oxygen (O
brated automatically.
If the calibration is successful, the detector beeps twice.
If the automatic O
calibration feature has been disabled,
2
Automatic O
span disabled displays.
2
10. The number of days remaining before calibration is due dis-
plays for all sensors.
16
) sensor is cali-
2
Note
If any sensor is over due for calibration, the LCD displays the
name of the sensor and the number of days past due.

Due-Lock Enabled

The Due-lock option ensures that a passcode must be entered when
calibration is past due, otherwise the detector automatically deactivates.
11. If Due-Lock is enabled in the user options, the following
screen displays.
Enter the correct passcode and press C to confirm.
If any sensor is overdue, Y displays continually until calibration
is performed.
Note