Setting the TWA Alarm Setpoint
The current TWA alarm setpoint displays for the selected sensor
(if applicable).
Press H or G to change the value for the TWA alarm setpoint. When
the required value displays, press C to confirm.
Setting the STEL Alarm Setpoint
The current STEL alarm setpoint displays for the selected sensor
(if applicable).
Press H or G to change the value for the STEL alarm setpoint. When
the required value displays, press C to confirm.
Calibration and Setting Alarm Setpoints
Setting the Low Alarm Setpoint
The current low alarm setpoint displays for the selected sensor.
Press H or G to change the value for the low alarm setpoint. When the
required value displays, press C to confirm.
Setting the High Alarm Setpoint
The current high alarm setpoint displays for the selected sensor.
Press H or G to change the value for the high alarm setpoint. When the
required value displays, press C to confirm.
Proceed to step #9
Setting the Remaining Alarm
GasAlertMicro 5/PID/IR
Setpoints.