K flashes as the detector initially detects the calibration gas.
GasAlertMicro 5 and PID
After 30 seconds the detector beeps and K stops flashing.
flashes while spanning the sensors until the detector
has attained a sufficient level of the expected gas.
Refer to
Table 13.
for times required to span.
Table 13. Time Required to Span
Gas Type
Most toxic gases and CO
2
Exotic toxic gases
LEL (combustibles)
PID gases
Insufficient Level: If a sensor does not attain a sufficient level
of expected gas, it is cleared from the LCD and is not spanned.
GasAlertMicro 5 IR
Time Required to Span
2 minutes
5 minutes
30 seconds
2 minutes
Calibration and Setting Alarm Setpoints
While the detector is spanning the sensor(s), a countdown of
time remaining displays in the lower left of the screen.
GasAlertMicro 5 and PID
When the span is complete, the following screen
displays.
Successful Span: Proceed to
Unsuccessful Span: Refer to Unsuccessful Span for possi-
ble solutions.
GasAlertMicro 5/PID/IR
GasAlertMicro 5 IR
Successful Span
step #6.