Asking For Repair - A&D FZ-120iWP Instruction Manual

Dust and waterproof precision balances fz-iwp series/fx-iwp series
Hide thumbs Also See for FZ-120iWP:
Table of Contents

Advertisement

Display
Other errors

17-3 Asking For Repair

If the balance needs service or repair, contact the local A&D dealer.
The balance is a precision instrument. Use much care when handling the balance and observe the
following when transporting the balance.
Use the original packing material.
Remove the weighing pan from the main unit
Error code
Description
Communications error
EC, E00
A protocol error occurred in communications.
Confirm the format, baud rate and parity.
Undefined command error
EC, E01
An undefined command was received.
Confirm the command.
Not ready
EC, E02
A received command can not be processed.
e.g. The balance received a Q command, but not in the
weighing mode.
e.g. The balance received a Q command while processing
a RE-ZERO command.
Adjust the delay time to transmit a command.
Timeout error
EC, E03
If the timeout parameter is set to "t-Up1", the balance did not
receive the next character of a command within the time limit
of one second.
Confirm the communication.
Excess characters error
EC, E04
The balance received excessive characters in a command.
Confirm the command.
Format error
EC, E06
A command includes incorrect data.
e.g. The data is numerically incorrect.
Confirm the command.
Parameter setting error
EC, E07
The received data exceeds the range that the balance can
accept.
Confirm the parameter range of the command.
If the errors described above can not be released or other
errors are displayed, contact the local A&D dealer.
65

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents