Response Adjustment; Automatic Response Adjustment - A&D BM-20 Manual

Analytical balance bm series
Hide thumbs Also See for BM-20:
Table of Contents

Advertisement

7. Response Adjustment

This function stabilizes the weight value by reducing the influence on weighing that is caused
by drafts and/or vibration at the place where the balance is installed. The adjustment of the
function can be performed by analyzing the environment automatically or hand-operation.
The state of the function has three stages as follows :
Indicator
Parameter Response
FAST
Cond 0
MID.
Cond 1
SLOW
Cond 2

Automatic Response Adjustment

7.1.
This function automatically updates the response adjustment by analyzing the influence of
the environment using the internal mass.
1 Press and hold the
displayed, then release the key.
2 The balance automatically sets the response characteristic.
Caution Do not allow vibration or drafts to affect the
balance during adjustment.
3 After automatic adjustment, the balance displays
returns to the weighing mode and displays the updated
response indicator for about thirty seconds.
Note
If the automatic response adjustment fails, the balance
displays
CH ng
breeze and vibration, also check the weighing pan. Then,
perform the adjustment again. Press the
return to the weighing mode.
If there is anything on the weighing pan, the balance
displays
. Remove the substance from the pan.
CH 0
Press the
CAL
Advise
If the automatic response adjustment is not helpful, try "7.2.
Manual Response Adjustment".
BM series
Fast response, Sensitive value
Slow response, Stable value
key until
MODE
. Check the ambient conditions such as
key to return to the weighing mode.
Response indicator
Stability
is
RESPONSE
,
end
key to
CAL
29
Press and hold
Release
Result

Hide quick links: