Oakton ORPTestr 50 Operating Instructions

Oakton ORPTestr 50 Operating Instructions

Pocket tester

Advertisement

Quick Links

Operating Instructions

ORPTestr
50 Pocket Tester
®
Applications
• Spas
• Aquaculture
• Food sectors
• Aquariums
• Hydroponics
• Studies
• Boilers
• Labs
• Water and wastewater
• Car washes
• Sanitation plants
• Water quality testing
• Ecology
• Steam generators
Getting Started
The ORPTestr 50 Pocket Tester has been factory calibrated and usually works
well out of the box. However, after extended periods of non-use, it is best to
remove the sensor cap and soak the sensor in warm tap water or ORP buffer
for 10 minutes or so. A brief rinse with deionized (DI) water is OK, but avoid
soaking or storing in deionized water as this will shorten ORP electrode life.
Prior to taking the measurements, periodic calibration with certified standards
is recommended for best accuracy.
Temperature Settings
1. Press ON/OFF ( ) to power on if needed.
v
2. Press MENU/
to enter setup window. Press HOLD/ to select Temp Set.
The display shows Set °C/°F and Temp Cal.
3. Press HOLD/ to select Set °C/°F. Display automatically shows °C and °F.
v
4. Scroll down by pressing MENU/
to toggle between °C and °F.
5. Press HOLD/ to select a temperature unit. The display shows the selected
temperature setting with a
.
Key Info
hold
hold
menu
Temp Set
Scroll
MENU
Set °C/°F
°C
Backlight
Accept
HOLD
Temp Cal
°F
Reset
EXIT
CAL
Backlight Settings
1. Press ON/OFF ( ) to power on if needed.
v
2. Press MENU/
to enter setup window. Scroll down by pressing
v
MENU/
to select Backlight. Press HOLD/ to select Settings. The
display shows Backlight.
3. Press HOLD/ to select Backlight. The display shows ON and OFF.
v
4. Scroll down by pressing MENU/
to toggle between ON and OFF.
Backlight ON increases readability in low-light conditions.
5. Press HOLD/ to select the desired backlight option. The display
shows the selected backlight option with a
.
Key Info
hold
menu
hold
Backlight
Scroll
MENU
Temp Set
Backlight
Backlight
ON
Accept
HOLD
OFF
Reset
EXIT
CAL
1065O100_MAN_35634-45.indd 1
ORP Calibration
Calibration should be done regularly, preferred once in a week. ORP tester
allows one point calibration.
1. Press ON/OFF ( ) to power on if needed.
2. Dip electrode in about 2 cm to 3 cm into the calibration standard solution.
3. Press the CAL/ESC key to enter calibration mode. The CAL indicator will be
displayed.
4. Allow the reading to stabilize. The timer icon blinks during the reading
stabilization. Once the reading stabilized the timer stops blinking.
5. Upper display shows current measured value based on the factory default.
6. Lower display shows current measured mV with last calibration (if no
treatment
calibration is performed earlier both display shows the same measurement
value).
in pools and more
v
7. Press MENU/
to adjust the mV value to match the calibration standard
value.
8. Press HOLD/ to confirm the calibration or press CAL/ESC key to exit the
calibration without confirmation.
cal
esc
ORP Measurement
1. Press ON/OFF ( ) to power on if needed.
2. Dip the electrode about 2 cm to 3 cm into the test solution. Stir and let the
reading stabilize. The timer icon blinks during this time. Once the reading is
stabilized, the timer stops blinking and
of the reading.
CAUTION: Testing dry samples is not accurate and can
lead to sensor damage or breakage. Soils must be wet
and free of particulates that may scratch the glass
sensor. Excessive force into dry samples can cause
hold
glass breakage.
°C
3. Note the ORP value or press HOLD/ to freeze the reading. To release the
reading, press HOLD/ again.
4. Press ON/OFF ( ) for 3 seconds to turn off tester. If a key is not pressed for
8.5 minutes, the tester will automatically shut off to conserve batteries.
For Sales & Service Contact
2650 E. 40th Ave. • Denver, CO 80205
Phone 303-320-4764 • Fax 303-322-7242
hold
OFF
www.geotechenv.com
CAL
CAL
CAL
Place the tester in
mV
RmV
the buffer to
189
844
start or press
cal
esc
189
to escape.
25.0°C
will appear to indicate the stability
ORP
RmV
844
25.0°C
1-800-833-7958
User Reset
Reset the ORP calibration to the default using the user reset function.
v
1. Press MENU/
to enter setup window. Scroll down by pressing MENU/
to select Reset. Press HOLD/ to select Reset. The display shows User
Reset and Factory Reset.
2. Press HOLD/ to select User Reset. The display automatically shows
No and Yes.
v
3. Scroll down by pressing MENU/
to toggle between No and Yes.
4. Press HOLD/ to confirm either No or Yes. The display shows the user
reset option with a
.
Key Info
hold
menu
hold
Temp Set
Scroll
MENU
User Reset
No
Settings
Accept
HOLD
Factory Reset
Yes
Reset
EXIT
CAL
Factory Reset
Reset the tester to the factory default using the factory reset function.
v
1. Press MENU/
to enter setup window. Scroll down by pressing the
v
MENU/
to select Reset. Press HOLD/ to select Reset. The display shows
User Reset and Factory Reset.
v
2. Scroll down by pressing the MENU/
to toggle between the resets.
Press HOLD/ to select Factory Reset. The display automatically shows
No and Yes.
v
3. Scroll down by pressing MENU/
to toggle between No and Yes.
4. Press HOLD/ to confirm either No or Yes. The display shows the factory
reset option with a
.
Key Info
menu
hold
hold
Temp Set
User Reset
Scroll
MENU
No
Settings
Accept
HOLD
Factory Reset
Yes
Reset
EXIT
CAL
Manual Temperature Setting
The MTS range is 0°C to 60.0°C (32.0°F to 140.0°F). User reset will set
temperature to default value 25°C or 77°F.
v
1. Press MENU/
to select setup window. Scroll down by pressing MENU/
to select Temp Set. Press HOLD/ to select Temp Set. The display shows
Set °C/°F and Temp Cal.
v
2. Scroll down by pressing MENU/
to toggle between Set °C/°F and
Temp Cal. Press HOLD/ to select Temp Cal.
3. The lower display shows the current measured temperature reading based
on the last set offset and the upper display shows the current measured
temperature reading based on factory default calibration.
4. Dip the tester into a solution of known temperature and allow time for the
built-in temperature sensor to stabilize.
v
5. Press MENU/
to adjust the temperature value or press the HOLD/ to
confirm the calibrated value as new temperature value of the solution.
Note: To exit without confirming the calibration, press CAL/ESC.
Key Info
hold
menu
hold
Temp Set
Scroll
MENU
Set °C/°F
Temp Cal
Backlight
23.4°C
Accept
HOLD
Temp Cal
23.4°C
Reset
EXIT
CAL
v
hold
hold
v
hold
23.4°C
9/5/2017 3:04:37 PM

Advertisement

loading

Summary of Contents for Oakton ORPTestr 50

  • Page 1: Operating Instructions

    8. Press HOLD/ to confirm the calibration or press CAL/ESC key to exit the MENU/ to select Reset. Press HOLD/ to select Reset. The display shows The ORPTestr 50 Pocket Tester has been factory calibrated and usually works calibration without confirmation. User Reset and Factory Reset.
  • Page 2 Prices are subject to change without notice. Accessories Ordering Code Product Description 35634-45 ORPTestr 50 pocket tester with case, lanyard, and batteries 35634-47 Replacement sensor module 35634-09 Replacement sensor cap 09376-00 Replacement alkaline batteries; AAA, 1.5 V. Pack of 12...