Temperature Measurements - Craftsman 82003 Owner's Manual

True rms multimeter
Hide thumbs Also See for 82003:
Table of Contents

Advertisement

TEMPERATURE MEASUREMENTS

1. Set the function switch to the ºF or ºC
position.
2. Insert the Temperature Probe into the
input jacks, making sure to observe
the correct polarity.
3. Touch the Temperature Probe head
to the part whose temperature you
wish to measure. Keep the probe
touching the part under test until the
reading stabilizes (about 30 seconds).
4. Read the temperature in the display.
Note: The temperature probe is fitted with a type K mini connector.
A mini connector to banana connector adaptor is supplied
for connection to the input banana jacks.
Note: The supplied thermocouple is rated for 482°F (250°C). For
higher temperatures you can purchase a high temperature
thermocouple from www.craftsman.com.
FREQUENCY SENSITIVITY (ELECTRICAL)
The frequency sensitivity is range dependent when the function
is selected from while in the voltage or current measuring
function. Below are typical sensitivities for the "electrical"
measurement modes.
Range
( DC/AC )
4V
40V, 400V
1000V/1000V
400mA
10A
Sensitivity
≥1.5V rms
≥10V rms
≥20V rms
≥420V rms
≥45mA rms
≥4A rms
19
Frequency width
5Hz~10kHz
5Hz~20kHz
5Hz~200kHz
50Hz~1kHz
5Hz~5kHz
5Hz~1kHz

Advertisement