EMC information

Guidance and Manufacturer's Declaration - Electromagnetic Emissions – This device is intended for use in the electromagnetic
environment specified below. The user of this device should make sure it is used in such an environment.
e
t
miSSionS
eSt
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/Flicker emissions
IEC 61000-3-3
Guidance and Manufacturer's Declaration - Electromagnetic Immunity – This device is intended for use in the electromagnetic
environment specified below. The user of this device should make sure it is used in such an environment.
i
t
mmunity
eSt
Electrostatic
Discharge (ESD)
IEC 61000-4-2
Electrical fast
Transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
IEC 61000-4-11
Power frequency (50/60 Hz)
magnetic field
IEC 61000-4-8
NOTE: U
is the a.c. mains voltage prior to application of the test level.
T
User Manual
c
ompliance
Group 1
The device uses RF energy only for its internal function. Therefore, its RF emissions
are very low and are not likely to cause any interference in nearby electronic
equipment.
Class B
The device is suitable for use in all establishments, including domestic establishments
and those directly connected to the public low-voltage power supply network.
Class A
Complies
iec 60601 t
eSt
l
evel
±6 kV contact
±8 kV air
±2 kV for power supply lines
±1 kV for input-output lines
±1 kV differential mode
±2 kV common mode
<5% U
T
(>95% dip in U
) for
T
0.5 cycle
40% U
T
(60% dip in U
) for
T
5 cycles
70% U
(30% dip in
T
U
) for 25 cycles
T
<5% U
(>95% dip in U
) for
T
T
5 sec
3 A/m
e
lectromagnetic
c
l
ompliance
evel
±6 kV contact
±8 kV air
±2 kV for supply mains
±1 kV for input/output lines
±1 kV differential mode
±2 kV for common mode
<5% U
T
(>95% dip in U
) for
T
0.5 cycle
40% U
T
(60% dip in U
) for 5 cycles
T
70% U
(30% dip in U
) for
T
T
25 cycles
<5% U
(>95% dip in U
) for
T
T
5 sec
3 A/m
e
- g
nvironment
uiDance
e
e
lectromagnetic
nvironment
g
uiDance
Floors should be wood, concrete or ceramic tile.
If floors are covered with synthetic material, the
relative humidity should be at least 30%.
Mains power quality should be that of a typical
home or hospital environment.
Mains power quality should be that of a typical
home or hospital environment.
Mains power quality should be that of a typical
home or hospital environment. If the user of
the device requires continued operation during
power mains interruptions, it is recommended
that the device be powered from an
uninterruptible power supply or a battery.
Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical hospital or home environment.
-