Table 5. Display Features (cont.)
Number
Feature
0L
--
Overload condition is detected.
Error Messages
bAtt
Replace the battery immediately.
In the capacitance function, too much electrical
di'c
charge is on the capacitor under test.
CAL Err
Invalid calibration data. Calibrate Product.
EEPr Err
Invalid EEPROM data. Have the Product serviced.
0PEn
Open thermocouple detected.
F2_
Invalid model. Have the Product serviced.
W Test lead alert. Shows when the test leads are in
A
mA/ A
the
or
terminal and the selected rotary
LEAd
switch position does not correspond to the terminal
being used.
Indication
True-rms Digital Multimeter

Automatic Power-Off

The Product automatically turns off if you do not turn the
rotary switch or push a button for 30 minutes. If MIN MAX
Recording mode is on, the Product will not turn off. Refer
to Table 6 to disable automatic power-off.
Input Alert™ Feature
If a test lead is connected to the mA/ A or A terminal, but
the rotary switch is not set to the correct current position,
the beeper warns you by making a chirping sound and
the display flashes "LEAd". This warning is intended to
stop you from attempting to measure voltage, continuity,
resistance, capacitance, or diode values with the leads
are plugged into a current terminal.
W Caution
To prevent damage, do not put the probes
across (in parallel with) a circuit with power
with a lead connected to a current terminal.
This can cause damage to a circuit with
power and blow the Product fuse. This can
occur because the resistance through the
current terminals of the Product is very low,
and causes a short circuit.