Table 7. Functions and Trigger Levels for Frequency Measurements
Function
Range
6 V, 60 V,
600 V, 1000 V
600 mV
600 mV
V
6 V
V
60 V
V
600 V
V
1000 V
All ranges
A
600 A, 6000 A
60 mA, 400 mA
A
6 A, 10 A
32
Approximate
Trigger Level
5 % of scale
Most signals.
30 mV
High-frequency 5 V logic signals. (The dc-coupling of the L function can
attenuate high-frequency logic signals, reducing their amplitude enough to
interfere with triggering.)
40 mV
Refer to the measurement guidelines given before this table.
1.7 V
5 V logic signals (TTL).
4 V
Automotive switching signals.
40 V
Refer to the measurement guidelines given before this table.
100 V
Frequency counter characteristics are not available or specified for these functions.
5 % of scale
AC current signals.
30 A , 300 A
Refer to the measurement guidelines given before this table.
3.0 mA , 30 mA
.30 A, 3.0 A