General Specifications - Fluke 1630 Technical Data

Fluke 1630 earth ground clamp meter datasheet
Hide thumbs Also See for 1630:

Advertisement

Stakeless testing system
The Fluke 1630 uses the Stakeless testing method, 
which eliminates the need to disconnect parallel 
ground rods and find suitable locations for 
placing auxiliary ground stakes. This saves time 
and enables users such as industrial and utility 
electricians, and field service electricians and 
contractors, to perform earth ground loop tests 
in locations where it is not possible to use other 
techniques, such as inside buildings or on power 
pylons.  With the Stakeless testing method, earth 
ground stakes are no longer necessary.  The Fluke 
1630 Earth Ground Clamp is placed around the 
earth ground rod or the connecting cable. A known 
voltage is induced by one half of the clamp and 
the current is measured by the other half. The 
tester automatically determines the ground loop 
resistance at this grounding connection. 

General specifications

Operating error
Storage temp. range
Reference temp. range
Storage humidity
Operating humidity
Display
Protection
Safety
Weight
Conductor size
Dimension (length x width x depth)
Emission
Immunity
Range selection
Overload indicator
Measuring time
Measuring frequency
Battery type
Battery life
Power consumption
Low battery indication
2  Fluke Corporation     Fluke 1630 Earth Ground Clamp Meter
Applications
  Earth Ground loop testing in various installations 
- for example, high voltage pylons, buildings, cell 
phone substations, RF transmitters, etc.
  Inspection of lightning protection systems
  Leakage current measurement
Refers to the operating temperature range and is guaranteed 
for 2 years
-20 °C to 60 °C (-4 °F to 140 °F ) 
23 °C ± 5 °C (73 °F ± 9 °F )
< 75 % R
H
< 85 % R
H
9999 digit LC-display with special symbols
IP20 according to IEC 529/EN 60529
300 V, CAT III / pollution degree 2 IEC 61010-1
640 g (1.4 lb) 
35 mm (1.38 in) approximately 
257 mm x 100 mm x 47 mm 
(1 in x 3.9 in x 1.9 in)
IEC 1000 4-2, IEC 61326-1 class B 
IEC 61000-4-2, 8 kV (air) criteria A
IEC 61000-4-3, 3 V/m performance criteria A
Auto
"OL" on display
0.4 sec
3.333 kHz
9 V alkaline (type IEC 6 LR 61)
≥ 8 hours (continuous operation)
40 mA
Display symbol