5.8 Water Trap / Filter Maintenance
The Water Trap / Flter Assembly removes water condensate from the gas
sample, and also prevents soot from contamnatng the nternal compo-
nents of the analyzer.
Drain the water condensate after every test.
Procedure:
1. Pull off the Inlet End Cap usng a slght twstng moton.
2. Pour out all water condensate; then reassemble trap.
Replace the Filter element when dirty.
Material Required: • Flter Element, P/N 07-1644
Procedure:
1. Pull off the Outlet End Cap usng a slght twstng moton.
2. Remove and dscard old flter.
3. Install new flter; then reassemble trap.
OUTLET
To replace Filter Element,
pull off Outlet End Cap using
a slight twisting motion.
42
FILTER ELEMENT
Figure 5-3. Water Trap / Filter Assembly
Fyrite Pro
To empty Water Trap,
pull off Inlet End Cap using
a slight twisting motion.
Instruction 24-9388
INLET