Summary of Contents for Bacharach Flue-Gas Sample Conditioning System
Page 1
Flue-Gas Sample Conditioning System P/N 0024-7224 Rev. 8 – May 2010 Introduction The Bacharach Flue-Gas Sample Conditioning System consists of the following components: • Probe with Peltier Cooler • Condensate Pump with Water Trap • Power Supply The function of this system is to remove water vapor from the fl ue-gas sample, thus preventing the formation of water droplets inside the probe hose.
100 and 240 VAC, 50/60 Hz. Connect the probe assembly’s stack-temperature thermocouple con- nec tor, pres sure fi tting, and fl ue-gas fi tting to the combustion an a lyz er per the analyzer’s instruction manual. Instruction 0024-9415 Bacharach, Inc.
Page 3
Flue Gas Sample Conditioning System Figure 2. Equipment Setup PROBE ASSEMBLY PELTIER COOLER Connect to Combustion Test Equipment STACK TEMP. THERMOCOUPLE PRESSURE (Blue) FLUE GAS (Yellow) PUMP 100 – 240 VAC 50/60 HZ POWER SUPPLY Bacharach, Inc. Instruction 0024-9415...
Operation When ready to conduct a combustion effi ciency test, plug in the power supply of the Flue-Gas Sample Conditioning System. You should hear the condensate pump start run ning and see the Peltier cooler’s red LED turn on and off as the cooler is cy cled. Note that the cool er’s operating time changes ac cord ing to its tem per a ture.
The pump’s wadding fi lter (Figure 3, Item 17) needs to be in spect ed at regular in ter vals for dirt contamination. Replace fi lter as nec es sary by fi rst unscrewing the fi lter chamber cap and then ex chang ing the fi lter with a new one. Bacharach, Inc. Instruction 0024-9415...
Filter Wadding (Qty. 1) ..........0024-1205 Plastic Cap w/ Drain ..........0106-5456 O-Rings: 9 x 2 mm ..............0105-5105 4 x 1.75 mm ..............0105-5106 50 x 3 mm ..............0105-5107 11 x 1 mm ..............0105-5108 Instruction 0024-9415 Bacharach, Inc.
Page 7
Flue Gas Sample Conditioning System Figure 3. Replacement Parts Part of Item 10 Part of Item 10 Bacharach, Inc. Instruction 0024-9415...
Flue Gas Sample Conditioning System Bacharach Sales/Service Centers Service can be obtained by contacting a Bacharach Service Center at the following locations: United States Bacharach, Inc. 621 Hunt Valley Circle New Kensington, PA 15068 Phone: 724-334-5051 Fax: 724-334-5723 Email: help@mybacharach.com Canada Bacharach of Canada, Inc.