THE ANALYZER WITH ITS FASTENING SCREWS.
ULTIMATELY, IT IS THE INSTALLER WHO ENSURES
THAT THE CONNECTIONS PROVIDE ADEQUATE
EMI/RFI SIELDING.
3.3.3.1 P
RIMARY
The power cord receptacle and fuse block are located in the same
assembly. Insert the power cord into the power cord receptacle.
CAUTION:
POWER IS APPLIED TO THE INSTRUMENT'S
CIRCUITRY AS LONG AS THE INSTRUMENT IS
CONNECTED TO THE POWER SOURCE.
The standard power supply requires 110 VAC, 50/60 Hz or 220
VAC, 50/60 Hz (optional) power.
3.3.3.2 F
I
USE
NSTALLATION
The fuse block, at the right of the power cord receptacle, accepts US
or European size fuses. A jumper replaces the fuse in whichever fuse
receptacle is not used.
3.3.3.3 50-P
IN
Figure 3-2 shows the pin layout of the Equipment Interface
connector. The arrangement is shown as seen when the viewer faces the
rear panel of the analyzer. The pin numbers for each input/output
function are given where each function is described in the paragraphs
below.
Figure 3-2: Equipment Interface Connector Pin Arrangement
3.3.3.4 A
NALOG
There are four DC output signal pins—two pins per output. For
polarity, see Table 3-1. The outputs are 0-1 VDC, 0-1 VDC, 4-20mA
DC, and 4-20mA DC.
rd
The 1st and 3
outputs (0-1 VDC, and 4-20mA DC) operate
analogously as do the 2
Teledyne Analytical Instruments
I
P
NPUT
OWER
E
I
QUIPMENT
NTERFACE
O
UTPUT
nd
th
and 4
outputs (0-1 VDC, and 4-20mA DC).
Installation
C
ONNECTOR
19