Setup: Negative O2 - Teledyne 3060E Instruction Manual

Ultra trace oxygen analyzer
Table of Contents

Advertisement

Model 3060E
Model 3060E
Model 3060E
Model 3060E
Model 3060E
NOTE: By choosing the external Zero/Span gas, whenever Zero or Span
calibration is initiated (Manual or Auto), the analyzer will remain
in the Analyze mode until Zero calibration is completed and
remain in the external Span mode until the Span calibration is
completed.
Setup: Negative O
Under normal operation conditions, the analyzer will display the O
tion from -12.5 ppb to 10 ppm (the negative value appears if the analyzer is
zeroed while the baseline is still trending downward). However, once the
analyzer has achieved base line stability and is zeroed, the analyzer will display
positive values. An exception to this is when the sample gas contains less than
1 ppb O
. Since the accuracy of the analyzer is ±1 ppb, with a sample gas
2
containing 0.5 ppb O
you prefer not to display the negative values, proceed with the following steps:
VALVES: ANALYZE
C=9:59
FLOW = 150.0 CC/M IN
5.00
PPM 02
AUTORANGE: 10 PPM
CAL
STDBY
SETUP
1
2
3
VALVES: ANALYZE
C=9:59
SELECT SYSTEM PASSW ORD SETUP
- - OR - -
SERIAL M ODEM , ZERO GAS SCRUB
OR NEGATIVE O2 READINGS
SYSTM
M ODEM
GAS
1
2
3
VALVES: ANALYZE
C=9:59
ENABLE NEGATIVE O2 VALUES?
NEGATIVE O2 CURRENTLY ENABLED
YES
1
2
3
4-56
6. To check your selection, press
again, confirm your selection and then
press
lyze mode main menu.
2
, the analyzer may display a signal of -.05 to 1.5 ppb. If
2
1. From the Analyze mode, press
T=25C
RANGE
ALRM R
4
5
2. Press
T=25C
NegO2
ESC
4
5
3. The display gives you the option of
T=25C
enabling or disabling negative O
ings. The last line indicates the present
selection. Once you press either
NO
ESC
, the system returns you to the screen
NO
4
5
in step 2. To check your selection, press
O2 again, and then press
NEG
to return to the Analyze mode.
Teledyne Electronic Technologies
Analytical Instruments
twice to return to the Ana-
ESC
O2.
NEG
Chapter 4
Chapter 4
Chapter 4
Chapter 4
Chapter 4
GAS
concentra-
2
.
SETUP
read-
2
or
YES
twice
ESC