Zero Calibration

THE SCREEN DISPLAYS WHETHER
THE ANALYZER IS USING THE
INTERNAL SCRUBBER
OR AN EXTERNAL ZERO GAS
PRESS ACK WHEN THE BACKGROUND
SIGNAL STABILIZES AND IS WITHIN
THE APPLICABLE LIMIT
NOTE:
If the background level does not
fall below the acceptable limit after
5-7 days, replace the electrolyte with
prepurged electrolyte. Refer to Sections
3.3.1 and 4.3.7.2 for more details
CALIBRATION PROCESS COMPLETED
PRESS ACK TO RETURN TO ANALYZE MODE
CAL
STDBY
SETUP
CAL
ZERO
MAN
Figure 6-6. Calibration/Zero Calibration
Teledyne Electronic Technologies
Analytical Instruments
Ultra T T T T T race Oxyg
Ultra
Ultra
race Oxyg
race Oxyg
en Anal
en Anal
Ultra
Ultra
race Oxygen Anal
race Oxyg
en Anal
en Analyz
RANGE
ALRMR
NOTE: Pressing ESC brings you
SCREEN WILL DISPLAY PRESENT
OXYGEN CONCENTRATION, FLOW RATE
AND TIME REMAINING
PRESSING ANY KEY WILL ABORT
AUTO SETTLING
PRESSING ACK AFTER
ABORTING AUTO SETTLING
RETURNS YOU TO ANALYZE MODE
AFTER AUTO SETTLING, THE SYSTEM
WILL AUTOMATICALLY RETURN
TO THE ANALYZE MODE
yz yz
yz yzer er er er er
back to previous screen.