Maintenance
2. To change between American and European fuses, remove
the single retaining screw, flip Fuse Block over 180 degrees,
and replace screw.
3. Replace fuse as shown in Figure 5-3.
4. Reassemble Housing as shown in Figure 5-2.
American Fuses

5.4 System Self Diagnostic Test

1. Press the System button to enter the system mode.
2. Use the __ arrow keys to move to More, and press Enter.
3. Use the __ arrow keys to move to Self-Test, and press Enter.
The following failure codes apply:

Table 5-1: Self-Test Failure Codes

Teledyne Analytical Instruments

Figure 5-3: Installing Fuses

Power
0
OK
1
5 V Failure
2
15 V Failure
3
Both Failed
Analog
0
OK
1
DAC A (0–1 V Concentration)
2
DAC B (0–1 V Range ID)
3
Both Failed
3000ZA
European Fuses
54