2640A/2645A NetDAQ
Users Manual
3. Verify Accuracy Configure the 5700A for the output values below and
verify the Spy window measurement is between the minimum and maximum
values. Change the channel 1 range as required (see Step 1).
Volts AC Range
300 mV
300 mV
300 mV
300 mV
3V
30V
300V [2640A only]
4. Close Spy Window.
Frequency Accuracy Test
Complete the following procedure to test the accuracy of the frequency function
for the 2640A and 2645A. Measurement accuracy applies to all channels, not just
the channel used for the test.
1. Configure Channel 1 for Frequency In NetDAQ Logger, configure channel
1 for frequency. There is no range selection for frequency as all frequency
measurements use Autoranging. (See "Configuring Analog Channel
Functions" in Chapter 3 of this manual.)
2. Open Spy Window Select the Spy command from the Utilities menu. Select
channel 0101 (instrument 01, channel 01) from the Channel list. Click OK to
open the Spy window.
3. Verify Accuracy Configure the 5700A for the output values below and
verify the Spy window measurement is between the minimum and maximum
values. Change the channel 1 range as required (see Step 1).
Frequency Range
Autorange Only
4. Close Spy Window.
6-14
5700A Output
20 mV@1 kHz
20 mV@100 kHZ
300 mV@1 kHz
300 mV@100 kHz
3V@1 kHz
30V@1 kHz
300V@1 kHz
5700A Output
1V@10 kHz
Minimum Reading
0.0197V
0.0185V
0.29885V
0.2845V
2.9885V
29.885V
298.85V
Minimum Reading
9.994 kHz
Maximum Reading
0.0203V
0.0215V
0.30115V
0.3155V
3.0115V
30.115V
301.15V
6-12.
Maximum Reading
10.006 kHz