3. Verify Accuracy Configure the 5700A for the output values below and
verify the Spy window measurement is between the minimum and maximum
values. Change the channel 1 range as required (see Step 1).
Volts DC Range
90 mV
90 mV
90 mV
300 mV
300 mV
300 mV
3V
3V
3V
30V
30V
30V
50V
50V
50V
4. Close Spy Window.

Volts AC Accuracy Test

Complete the following procedure to test the accuracy of the volts ac function for
both the 2640A and 2645A. Measurement accuracy applies to all channels, not
just the channel used for the test.
1. Configure Channel 1 for Volts AC In NetDAQ Logger, configure channel 1
for volts ac, 300 mV range. (See "Configuring Analog Channel Functions" in
Chapter 3 of this manual.)
2. Open Spy Window Select the Spy command from the Utilities menu. Select
channel 0101 (instrument 01, channel 01) from the Channel list. Click OK to
open the Spy window.
5700A Output
Short Circuit (Zero)
+90 mV
-90 mV
Short Circuit (Zero)
+300 mV
-300 mV
Short Circuit (Zero)
+3V
-3V
Short Circuit (Zero)
+30V
-30V
Short Circuit (Zero)
+50V
-50V
Minimum Reading
-0.000023V
+0.089965V
-0.090035V
-0.00005V
+0.29991V
-0.30009V
-0.0004V
+2.9992V
-3.0008V
-0.005V
+29.991V
-30.009V
-0.04V
+49.95V
-50.05V
Maintenance
Performance Test
Maximum Reading
+0.000023V
+0.090035V
-0.089965V
+0.00005V
+0.30009V
-0.29991V
+0.0004V
+3.0008V
-2.9992V
+0.005V
+30.009V
-29.991V
+0.04V
+50.05V
-49.95V
6-11.
6
6-13