B-1. 2645A Crosstalk Specifications - Fluke NetDAQ 2640A User Manual

Networked data acquisition unit
Hide thumbs Also See for NetDAQ 2640A:
Table of Contents

Advertisement

the routing of your input wiring: keep it away from or shielded from all ac
power mains wiring and other sources of noise.
3. Channel Spacing When possible, leave as many unused channels as possible
between your ac voltage channel inputs and your low level inputs.
Crosstalk from a VAC or frequency signal on a 2645A channel to an adjacent
channel using the Slow reading rate is as shown in Table B-1:
VAC/Frequency
Volts DC
<100 Hz
100 Hz
500 Hz
1 kHz
2 kHz
5 kHz
10 kHz
20 kHz
50 kHz
100 kHz
>100 kHz
Note 1 The temperature ranges are particularly susceptible to crosstalk at frequencies above 20
kHz, and that the error goes up as the square of the voltage, e.g., 2V ac at 100 kHz can cause a
4 degree error on a thermocouple. Because of this sensitivity, it is not advisable to mix
temperature signals and high frequency signals in the same 2645A.
Noise, Shielding, and Crosstalk Considerations
Table B-1. 2645A Crosstalk Specifications.
Resistance
-80 dB
.2%/V if R>3k
-100 dB
.1%/V if R>3k
-100 dB
.1%/V if R>3k
-50 dB
.1%/V if R>3k
-60 dB
.05%/V if R>3k
-66 dB
.1%/V if R>3k
-80 dB
.01%/V
-50 dB
.02%/V
-40 dB
.05%/V
-40 dB
.1%/V
-40 dB
.1%/V
Crosstalk Considerations
Temperature
Temperature
(100 Ohm RTD)
(Thermocouple)
.04°/V
.02°/V
.02°/V
.02°/V
.02°/V
.02°/V
.02°/V
.05°/V
.1°/V
.2°/V
.4°/V
B
.1°/V
.003°/V
.03°/V
.3°/V
.1°/V
.1°/V
.1°/V
.1°/V
.5°/V
1°/V
2°/V
B-3