Front Panel Operating Procedures; Power-On Options; Instrument Default Parameters - Fluke NetDAQ 2640A User Manual

Networked data acquisition unit
Hide thumbs Also See for NetDAQ 2640A:
Table of Contents

Advertisement

Front Panel Operating Procedures

Power-On Options

There are three power-on options as listed below:
Normal Power-On Turn power switch on. The instrument communication
parameters are the same as when the instrument was last turned off.
Configuration-Reset Power-On Hold the front panel COMM key down,
and then turn the power switch on. Continue holding the COMM key until the
instrument beeps. The communication parameters are reset to default values
(see Table 2-3).
Display-Hold Power-On Hold the front panel arrow-left ( < ) key down, and
then turn the power switch on. Continue holding the < key until the instrument
beeps. The instrument front panel display remains on until you press any front
panel key. This allows inspection of the display segments.
Each power-on sequence includes a four-second self-test routine. If the self-test
fails, the instrument displays ERROR in the primary display with a code character
in the secondary display. See Chapter 6 "Maintenance" for information on error
codes. Power-on also clears channel configuration data and sets all channels to
OFF.
Parameter
Base Channel Number
Line Frequency
Network Selection
Socket Port
Internet Protocol Address
Baud Rate
Default Gateway
Table 2-3. Instrument Default Parameters
Preparing for Operation
Instrument Preparation
Default Setting
1
60 Hz
Isolated Network
4369
---.---.---.--- (dashes)
19200
OFF
2
2-25.
2-26.
2-21