Page 2
175, 177 & 179 Users Manual Supplement Change #1, 36861, 38746 On page 2, under Terminals, replace Item 1 & 2 with the following: Input terminal for AC and DC milliamp measurements to 400 mA and frequency measurements. Input terminal for AC and DC current measurements to 10 A and frequency measurements.
Page 3
Manual Supplement 175, 177 & 179 Users On page 4, under MIN MAX AVG Recording Mode, replace the Note with following: Note For DC functions, accuracy is the specified accuracy of the measurement function ± 12 counts for changes longer than 350 ms in duration.
Page 4
175, 177 & 179 Users Manual Supplement On page 6, under Measuring AC and DC Voltage, replace the figure with the following: Volts AC Volts DC Millivolts DC HOLD MIN MAX RANGE HOLD MIN MAX RANGE HOLD MIN MAX RANGE AIK03F.eps...
Page 5
Manual Supplement 175, 177 & 179 Users On pages 11-14, replace the entire Specifications with the following: Specifications Accuracy is specified for 1 yr after calibration, at operating temperatures of 18 °C to 28 °C, with relative humidity at 0 % to 90 %. Accuracy specifications take the form of: ±...
Page 6
175, 177 & 179 Users Manual Supplement Accuracy ± ( [ % of Reading ] + [ Counts ] ) Function Range Resolution Model 175 Model 177 Model 179 0.1 mV 1.0 % + 3 1.0 % + 3 1.0 % + 3 600.0 mV...
Page 7
Manual Supplement 175, 177 & 179 Users Accuracy ± ( [ % of Reading ] + [ Counts ] ) Function Range Resolution Model 175 Model 177 Model 179 60.00 mA 0.01 mA 400.0 mA 0.1 mA DC Amps 1.0 % + 3 1.0 % + 3...