Pushbutton Functions - Fluke 705 Instruction Sheet

Loop calibrator
Hide thumbs Also See for 705:
Table of Contents

Advertisement

Pushbutton Functions

Button
M
M
P
+
P
D
U
+
D
U
or
D
U
Using the Current Output Modes
The calibrator provides current output for calibrating and testing
0 to 20 mA and 4 to 20 mA current loops and instruments.
In source mode, the calibrator supplies the current. In simulate
mode, the calibrator simulates a two-wire transmitter in an
externally-powered current loop.
Press to select source, simulate, or measure
modes.
Press while turning on the calibrator to select loop
power. The Loop symbol appears in the display.
Press simultaneously to start the SpanCheck
S
function. The
display. Output toggles between 0 % and 100 %.
Press to manually step the output up or down 25
%.
Press both buttons simultaneously to select auto
ramp output. The
display.
Press to scroll the display up or down.
Press while turning on the calibrator to change and
save scroll steps to 0.001 mA.
Press while turning on the calibrator to toggle
between 0 to 20 mA and 4 to 20 mA and save the
current output span in a nonvolatile memory.
Function
symbol appears in the
EON
symbols appear in the