187/189
Calibration Manual
Testing Voltage, Current, Resistance, Capacitance and Diode Functions
To verify the accuracy of Meter functions, do the following.
1. Connect the Calibrator to the
2. Turn the rotary switch to the function listed in each step.
3. Apply the input from the steps listed in Table 4.
4. Compare the reading on the Meter display with the display reading in Table 4.
5. If the display reading falls outside of the range shown in Table 4, the Meter does not
6. Repeat steps 2 through 4 for the remaining inputs shown in Table 4.
Step
Function
1
AC V
2
AC V
3
AC V
4
AC V
5
AC V
6
AC V
7
AC V
8
AC V
9
AC mV
10
AC mV
11
AC mV
12
AC mV
13
AC mV
14
AC mV
15
Frequency
16
Duty Cycle
17
DC V
18
DC V
19
DC V
20
DC V
21
DC V
22
DC V
1.
Do not rely on autoranging for all tests. Some test steps will require manual ranging.
2.
Must allow for source and meter settling before recording reading.
3.
Wait at least 60 seconds for source and meter to settle before recording reading.
Using Relative Mode (REL) to zero residual reading (apply 0 Ω and allow UUT to settle for 45 seconds before
4.
pushing REL).
18
meet specification.
Table 4. Performance Test Steps
1
Range
5.0000 V
5.0000 V
5.0000 V
50.000 V
50.000 V
50.000 V
500.00 V
1000.0 V
50.000
50.000
50.000
500.00 mV
3000.0 mV
3000.0 mV
50.000 kHz
3.0000 V
5.0000 V
5.0000 V
5.0000 V
50.000 V
500.00 V
1000.0 V
z
and
COM
inputs on the Meter.
Frequency or
Input Level
Model
0.25 V
100 kHz
5 V
20 kHz
5 V
100 kHz
50 V
700 Hz
50 V
20 kHz
50 V
100 kHz
500 V
65 Hz
500 V
65 Hz
2.5 mV
100 kHz
50 mV
20 kHz
50 mV
100 kHz
500 mV
10 kHz
3 V
20 kHz
3 V
100 kHz
1 V
20 kHz
2.2 Vp-p,
500 Hz
@30%
Sq. wave
short
5 V
-5 V
-50 V
-500 V
-500 V
2
Display
Lower
Upper
Limit
Limit
0.2260
0.2740
4.9210
5.0790
4.5960
5.4040
49.760
50.240
49.210
50.790
45.960
54.040
497.60
502.40
494.0
506.0
2.085
2.915
46.850
53.150
42.460
57.540
474.60
525.40
2951.0
3049.0
2756.0
3244.0
19.998
20.002
21.82
38.18
-0.0010
0.0010
4.9977
5.0023
-4.9977
-5.0023
-49.982
-50.018
-499.48
-500.52
-499.3
-500.7