Fluke III Series Instruction Sheet

Fluke III Series Instruction Sheet

Hide thumbs Also See for III Series:

Advertisement

70/73 English Instruction Sheet
Series III Multimeter
W Read First: Safety Information
To ensure that the meter is used safely, follow these instructions:
• Never use the meter if the meter or test leads look damaged.
• Be sure the test leads and rotary switch are in the correct
position for the desired measurement.
• Never measure resistance in a circuit when power is applied.
• Never touch the probes to a voltage source when the test
leads are plugged into the 10 A or 300 mA input jack.
• To avoid damage or injury, never use the meter on circuits that
exceed 4800 volt-amps.
• Never apply more than the rated voltage between any input
jack and earth ground.
• Be careful when working with voltages above 60 V dc or 30 V
ac rms. Such voltages pose a shock hazard.
• Keep your fingers behind the finger guards on the test probes
when making measurements.
To avoid false readings, which could lead to
possible electric shock or personal injury, replace
the battery as soon as the battery indicator (N)
appears.
Symbols
W
Read First: Safety Information
Z
Manual Range or Automatic Touch Hold
T
Double Insulation
CAT II
Overvoltage Installation Category per IEC 1010-1.
Typical CAT II locations include main wall outlets,
local appliances, and portable equipment.
PN 650454
August 1997 Rev.2, 2/98
© 1997, 1998 Fluke Corporation. All rights reserved. Printed in U.S.A.
All product names are trademarks of their respective companies.
Page 1
70/73
Instruction Sheet
Warning
®
mode
®

Advertisement

Table of Contents
loading

Summary of Contents for Fluke III Series

  • Page 1 Overvoltage Installation Category per IEC 1010-1. Typical CAT II locations include main wall outlets, local appliances, and portable equipment. PN 650454 August 1997 Rev.2, 2/98 © 1997, 1998 Fluke Corporation. All rights reserved. Printed in U.S.A. All product names are trademarks of their respective companies.
  • Page 2: Input Jacks

    70/73 English Instruction Sheet Page 2 Input Jacks Amps Volts, Ohms, Diode Test 600V FUSED Common Terminal Milliamps hr2f.eps See Specifications for overload protection. Pushbutton Use the pushbutton to either select a fixed range or put the meter in the Automatic Touch Hold mode.
  • Page 3: Bar Graph

    70/73 English Instruction Sheet Page 3 Automatic Touch Hold Mode ® W Warning ® To avoid electric shock, do not use the Touch Hold to determine if a circuit with high voltage is dead. ® The Touch Hold mode will not capture unstable or noisy readings.
  • Page 4 70/73 English Instruction Sheet Page 4 Standby If the meter is on but is inactive for an hour (20 minutes in diode test), the screen goes blank and displays selected bar graph segments. To resume operation, turn the rotary switch or press AC and DC Voltage (K L m L) Volts AC Volts DC...
  • Page 5: Diode Test (G)

    70/73 English Instruction Sheet Page 5 Diode Test (G) Good Diode Good Diode Single Beep Forward Bias Reverse Bias Bad Diode Bad Diode Open Shorted hr7f.eps...
  • Page 6 70/73 English Instruction Sheet Page 6 Continuity Test ( R ) hr5f.eps If continuity exists (resistance < 210 Ω), the beeper sounds ® continuously. The meter beeps twice if it is in the Touch Hold mode. Current (? A) W Warning To avoid injury, do not attempt a current measurement if the open circuit voltage is above the rated voltage of the meter.
  • Page 7: Maintenance

    70/73 English Instruction Sheet Page 7 Holster The snap-on holster absorbs shocks and protects the meter and comes with a Flex-Stand™. To protect the front of the meter, put the meter face down in the holster. hr15f.eps Maintenance W Warning To avoid electric shock, remove the test leads before opening the case, and close the case before using the meter.
  • Page 8 70/73 English Instruction Sheet Page 8 Internal Fuse Test 10 A 300 mA Ω Ω <0.5 hr10f.eps Battery and Fuse Replacement Note Before opening the case, make sure the test leads are removed and the rotary switch is turned to OFF. hr11f.eps...
  • Page 9: Specifications

    To clean the meter, use a damp cloth and mild detergent; do not use abrasives or solvents on the meter. Service and Parts For service information in the U.S.A. and Canada, call 1-800-44- FLUKE (1-800-443-5853), in Europe, 31 40 267 8200, from other countries, 1-425-356-5500. Item Description Fluke Quan.
  • Page 10 70/73 English Instruction Sheet Page 10 Electromagnetic Compatibility Total accuracy = Specified accuracy in RF field of 3 V/m on all plus +2.0% of range for all functions functions except MΩ except: 320.0 mV range: total accuracy = specified accuracy + 1.0% of range 320.0Ω...
  • Page 11 70/73 English Instruction Sheet Page 11 Function Range Accuracy ±(0.3%+1) 3.200 V, 32.00 V, 320.0 V ±(0.4%+1) 600 V ±(0.3%+1) 320.0 mV ±(2%+2) (45 to 500 Hz, 3.2V 3.200 V, 32.00 V, ±(2%+2) range. Other ranges 45 320.0 V, 600 V to 1 kHz) 320.0 Ω...
  • Page 12 70/73 English Instruction Sheet Page 12 Overload protection for all functions and ranges: 600 V. Function Input Impedance (Nominal) L, m L, K >10 MΩ, <50 pF Common Mode Normal Mode Rejection Ratio (1 Rejection kΩ Unbalanced) L, m L >120 dB at dc, 50 Hz, >60 dB at 50 Hz or 60 Hz or 60 Hz...
  • Page 13 Test Accessories - Other Click to view products by manufacturer: Fluke Other Similar products are found below : AO1026 AO1035 AO1051 AO1059 149118 C20 RS40 RS41 AO1030 AO1032 AO1036 AO1058 4742 KTS288540 00415 00163 046450 GPR10 ADPTR-HDV VS-AVT-C02-L03 VS-AVT-CABLE-04 VS-AVT-CABLE-16 VS-AVT-CABLE-20 VS-AVT-CABLE-30 VS-CKP14-6 VS-CKP4-000 VS-AVT2-C02L03 VS-AVT2-C08L10 CKSB1-00 CKSB6-2 CKSB10-8 GP73800080 MPT-VF 98-CAL020 LSP08-RTS 1-1437358-1 4408 POMONA AX-904 DME D110 T1 LMLH 50 1508-POM 00836 85392-12 207P18.2222.F CR-31 7708...