5 220A
Instruction Manual
selection switches are inside the 5220A. Refer to Section 4 of this manual for information
on how to properly set the line power switches.
Accessories available for use with the 5220A are listed and described in Table 1-1. The
rack mounting kits are designed for use with a standard 19-inch equipment rack. The
Model Y5020A is a precision current shunt recommended for use in calibrating the
5220A. Specify both model number and description when ordering accessories.
1-2. Specifications
Specifications for the Model 5220A are given in Table 1-2.
Calibration Cycle
Basic
Transconductance
Output Range
Compliance Voltage
DC Accuracy
AC Accuracy
Short Term DC Stability
Short Term AC Stability
Harmonic Distortion and Noise
Temperature Coefficient
Transient Recovery
1-2
Table 1-1. Accessories
Model No.
M07-205-600
M00-260-610
M00-280-610
Y5020
Table. 1-2. Specifications
All specifications are valid for a 180-day period at an operating
temperature of 20 C to 30 C and a relative humidity of 70 % or
less.
1 Siemen (1 ampere per volt)
0 A to 20 A dc or rms ac (28.3 A peak)
4 V dc or 3 V rms ac (4.25 V peak)
(0.025 % of output +1 mA)
(0.05 % of output + 1 mA) from 30 Hz to 1 kHz, and (0.05 % of
output + 1 mA) x f from 1 kHz to 5 kHz, where f = frequency in kHz.
Output changes less than (0.005 % + 200 A) in 10 minutes, with
constant line, load, and temperature.
Output changes less than (0.01 % + 500 A) in 10 minutes, with
constant line, load, and temperature.
(0.05 % of output 1 mA rms) over frequency range of 30 Hz to 1
kHz and measured with a noise bandwidth of 300 kHz, (0.05 % of
output + 1 mA) x f from 1 kHz to 5 kHz, where f = frequency in kHz.
(0.0025 % of output + 100 A) per degree C above 30 C or
below 20 C.
Output will settle to within 0.01 % of final value within 2 seconds
following a programmed change in output current or frequency
(10 ms for 5220A alone).
Description
Rack Mounting Kit
Rack Slide Kit, 18 in
Rack Slide Kit, 24 in
Precision Shunt