4-1. Introduction

These servicing instructions are for use by qualified personnel
only. To avoid electric shock, do not perform any servicing
other than that contained in the operation instructions unless
you are qualified to do so.
This section of the manual contains maintenance information for the Model 5220A
Transconductance Amplifier. The material is presented under the categories of shipping
information, general maintenance, performance test, calibration adjustments, and
troubleshooting. The performance test is recommended as an acceptance test when the
unit is first received, and later as a calibration procedure to verify the 180-day
specifications. Table 4-1 lists the equipment required for both the performance test and
calibration adjustments. If a recommended model is not available, an equivalent
instrument meeting the required characteristics can be substituted.
Equipment Type
DC and AC Voltage Calibrator
Digital Multimeter
Precision Shunt
Load Resistor
Load Resistor
Distortion Analyzer
Autotransformer
WWarning
Table 4-1. Required Test Equipment
Required Specifications
0 to 20 V dc 0.002 %
0 to 20 V dc 0.002 %, Resolution: 1 V
0.01
0.01 % dc, 0.035 % ac to 5 kHz
0.01
3 %, 50 W
0.01
5 %, 10 W
Frequency range: 100 Hz to 5 kHz
Nominal line voltage 10 %, 500 W
Section 4

Maintenance

Recommended Type
Fluke Model 5700A
HP3458A
Fluke Model Y5020
Dale RH-50
Dale RH-10
Sound Technology Model
1700A
General Radio Variac
4-1