General Specifications - Fluke 5725A Getting Started

Hide thumbs Also See for 5725A:
Table of Contents

Advertisement

5725A
Amplifier

General Specifications

Warm-Up time: 2 X the time since last warmed up, to a
maximum of 30 minutes.
System installation: Rear output configuration and
rack- mount kit available.
Standard interfaces: IEEE-488, RS-232, 5725A,
5205A or 5215A, 5220A, phase lock in (BNC), phase
reference out (BNC).
Temperature performance: Operating: 0 °C to 50 °C.
Calibration: 15 °C to 35 °C. Storage: -40 °C to 75 °C.
Relative humidity: Operating: <80 % to 30 °C, <70 %
to 40 °C, <40 % to 50 °C.
Storage: <95 %, non-condensing.
Safety: Designed to comply with UL1244 (1987); IEC
348-1978; IEC 66E (CO) 4; CSA 556B.
Guard isolation: 20 volts
EMI/RFI: Designed to comply with FCC Rules Part 15,
Subpart J, Class B; VDE 0871, Class B.
Reliability: MIL-T-28800D, para. 3.13.3.
6
Line Power: 47 to 63 Hz; ±10 % allowed about
selectable nominal line voltage: 100V, 110V, 115V,
120V, 200V, 220V, 230V, 240V. Maximum power:
5700A, 300 VA; 5725A, 750 VA.
Electromagnetic Compatibility: This instrument is
designed to operate in Standards Laboratory
environments where in electromagnetic is highly
controlled. If used in areas with RF fields >0.78 V/m there
could be errors in measurements.
Size:
5700A: Height 17.8 cm (7 in), standard rack increment,
plus 1.5 cm (0.6 in) for feet; Width 43.2 cm (17 in),
standard rack width; Depth 63.0 cm (24.8 in), overall;
57.8 cm (22.7 in), rack depth.
5725A: Height 13.3 cm, (5.25 in); Width and depth same
as 5700A. Both units project 5.1 cm, (2 in) from rack
front.
Weight: 5700A: 27 kg (62 lbs); 5725A: 32 kg (70 lbs).